- Electron beam dosimetry
- Commissioning of both radiation therapy planning systems and radiation therapy devices
- Complies with the design principles outlined by Dr. M. Rus
- Meets AAPM (American Association of Medical Physicists) TG-51 and IAEA TRS-398 for low energy beams
- Meets the highest performance requirements (IEC 60731)
- Good shielding reduces backscatter effects
- Ventilation provides air density correction and eliminates the need for a radioactive stability checker
- White camera body allows easy installation with respect to crosshairs and lasers
Effective point (mm): 1.07 below the surface of the entrance window
Active area height (mm): 2,0
Active area diameter (mm): 15,6
Protective ring width (mm): 4,1
Polarization effect: Within 1,000 (±) 0.01
Maximum dose rate for (Gy / s):
≥ 99.5% Saturation 5.2
≥ 99.0% Saturation 10.4
Maximum pulse dose for (mGy):
≥ 99.5% Saturation 0.46
≥ 99.0% Saturation 0.92
Radiation quality: Photons: 60Co up to 25 MV Electrons: 5 MeV - 25 MeV
Connector: TNC or BNC