Features:
- Adjustable flow rate, ml / min - 0.1-30;
- maximum working pressure, bar - 50;
- possibility of simultaneous connection of up to 16 buffer solutions;
- optical module UV, nm - 260/280/300;
- measured range of concentration of hydrogen ions, units. pH - 1-14;
- measured range of electrical conductivity, mS / cm - 1-150;
- dimensions, W × D × H, mm - 509 × 450 × 584;
- weight, kg - 47.
The main stages of planning and carrying out the MCSGP procedure for the chromatographic separation of complex multicomponent mixtures:
- identify the peak of the investigated product according to the results of the preliminary chromatographic procedure and save the obtained chromatogram in the database;
- using the Contichrom software setup wizard, the boundaries of the peak of the pure target product and the gray zone, including the area of its contamination with foreign impurities, are determined. Then designate and fix in the taskbar of the chromatograph the start and end points of the recycles of the purification of fractions;
- start the process of carrying out automatic fine purification of the target product using MCSGP.